Page 1 - Technical Service Manual

For use with List Numbers 12348 and 12618Technical Service ManualHospira, Inc.275 North Field DriveLake Forest, IL 60045 USATM 430-95424-009 (A, 201

Page 2

CONTENTSPlum A+3 Infusion SystemviiiTechnical Service ManualFigure 5-40 Pressing the Pumping Chamber . . . . . . . . . . . . . . . . . . . . . . . .

Page 3 - ChangeHistory

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 38 Technical Service Manual6. Press in firmly on the pumping chamber to pump air o

Page 4 - CHANGE HISTORY

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 39 Plum A+3 Infusion System5.3.2.5DISTAL OCCLUSION TEST SETUPThis section describes the

Page 5 - Contents

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 40 Technical Service Manual5.3.2.5.2Setup ProcedureThe three-way stopcock has two

Page 6 - SECTION 5

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 41 Plum A+3 Infusion System4. Place the DPM connector at a height of 0  12 inches from

Page 7 - SECTION 7

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 42 Technical Service Manual5.3.3SELF TESTThe self test procedure uses the Basic te

Page 8 - SECTION 9

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 43 Plum A+3 Infusion System5.3.5UNRESTRICTED FLOW TESTTo perform the unrestricted flow t

Page 9

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 44 Technical Service Manual5.3.7KEYPAD VERIFICATION/FUNCTIONAL TESTTo perform the

Page 10 - Plum A+3 Infusion System

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 45 Plum A+3 Infusion System5.3.8ALARM LOUDNESS TESTTo perform the alarm loudness test, u

Page 11

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 46 Technical Service Manual5.3.9KEYPAD LOCKOUT SWITCH TESTTo perform the keypad lo

Page 12

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 47 Plum A+3 Infusion System5.3.10PROXIMAL OCCLUSION TESTTo perform the Proximal Occlusio

Page 13 - INTRODUCTION

CONTENTSTechnical Service ManualixPlum A+3 Infusion SystemTablesTable 1-1 Conventions . . . . . . . . . . . . . .

Page 14 - COMPONENT DESIGNATORS

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 48 Technical Service Manual5.3.12DISTAL AIR-IN-LINE TESTThe Distal Air-in-Line tes

Page 15 - ACRONYMS AND ABBREVIATIONS

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 49 Plum A+3 Infusion System11. Position the control arm of the three-way stopcock over t

Page 16 - SECTION 1 INTRODUCTION

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 50 Technical Service Manual16. Open the three-way stopcock to air by positioning t

Page 17 - INFUSION SYSTEM INSTALLATION

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 51 Plum A+3 Infusion System24. Set the three-way stopcock to measure pressure by positio

Page 18 - SELFTEST

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 52 Technical Service ManualFigure 5-54. Delivery Accuracy Test Setup2. Turn on the

Page 19 - HOSPIRA Plum A+

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 53 Plum A+3 Infusion System11. Enter a rate of 200 mL/hr and a VTBI of 10 mL.12. Press [

Page 20 - BIOMED SETTINGS

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 54 Technical Service Manual3. Turn on the DMM and set it to measure resistance.4.

Page 21

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 55 Plum A+3 Infusion System5.3.17END OF THE PVTIf any tests fail, see Section 6, or cont

Page 22

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 56 Technical Service ManualThis page intentionally left blank.

Page 23

Technical Service Manual 6 - 1 Plum A+3 Infusion SystemSection6TROUBLESHOOTINGThis section contains information on technical assistance, warning mess

Page 24 - ALARMSLOG

CONTENTSPlum A+3 Infusion SystemxTechnical Service ManualThis page intentionally left blank.

Page 25

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 2 Technical Service Manual6.3ALARMMESSAGESANDERRORCODESUnder most alarm conditions the infus

Page 26

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 3 Plum A+3 Infusion SystemTable 6-2. Operational Alarm Messages and Corrective ActionsA

Page 27 - WARRANTY

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 4 Technical Service ManualN182orE182Prox. Occl B, Air or Prox. Occl BNegative proximal occlusion

Page 28 - SECTION 2 WARRANTY

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 5 Plum A+3 Infusion SystemN189orE189Prox. Occl B, Air Peak proximal occlusion B, delive

Page 29 - SYSTEMOPERATINGMANUAL

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 6 Technical Service ManualN233orE233Distal air cumulativeDistal air cumulative500 mL of air dete

Page 30

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 7 Plum A+3 Infusion SystemN255 Lockout violation Soft lockout violationThe use of the [

Page 31 - THEORYOFOPERATION

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 8 Technical Service Manual6.3.2ERRORCODESREQUIRINGTECHNICALSERVICETable 6-3 lists infusion s

Page 32 - ELECTRONICSUBSYSTEMOVERVIEW

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 9 Plum A+3 Infusion SystemE321 Failure of charging current to drop below end of charge

Page 33 - CPUSUBSYSTEM

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 10 Technical Service ManualE326 Battery disconnected while the infuser is on or battery voltage

Page 34 - LCD CONTROLLER

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 11 Plum A+3 Infusion SystemE378 I/O valvephase lossGeneric I/O valve failureTurn power

Page 35 - LCD BACKLIGHT CONTROL

Technical Service Manual 1 - 1 Plum A+3 Infusion SystemSection1INTRODUCTIONThe Plum A+™ 3 infusion system is designed to meet the growing demand for

Page 36 - ANALOG-TO-DIGITAL CONVERTER

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 12 Technical Service ManualE436 ROM failure ROM checksum failure Turn power off, then on,to rese

Page 37

6.3 ALARM MESSAGES AND ERROR CODESTechnical Service Manual 6 - 13 Plum A+3 Infusion SystemE448 SEEP write failureSEEP data write failed Replace mechan

Page 38 - KEYPAD LOCKOUT INTERFACE

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 14 Technical Service ManualThe following error codes are not generated in the Biomed service mod

Page 39 - DATAPORT INTERFACE

6.4 TROUBLESHOOTING PROCEDURESTechnical Service Manual 6 - 15 Plum A+3 Infusion System6.4TROUBLESHOOTINGPROCEDURESThis section details recommended pr

Page 40 - POWER SUPPLY INTERFACE

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 16 Technical Service ManualProximal Occlusion TestSection 5.3.10Closed proximal clamp Open clamp

Page 41 - MECHANISM INTERFACE

6.4 TROUBLESHOOTING PROCEDURESTechnical Service Manual 6 - 17 Plum A+3 Infusion System6.4.1UNRESTRICTED FLOWUnrestricted flow may occur if the infuser

Page 42 - SECTION 4 THEORY OF OPERATION

SECTION 6 TROUBLESHOOTINGPlum A+3 Infusion System 6 - 18 Technical Service ManualThis page intentionally left blank.

Page 43 - POWERSUPPLYSUBSYSTEM

Technical Service Manual 7 - 1 Plum A+3 Infusion SystemSection7REPLACEABLEPARTSAND REPAIRSThis section itemizes all parts and subassemblies of the

Page 44 - POWER CONTROL

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 2 Technical Service Manual7.2.2REQUIRED TOOLS AND MATERIALSThe following tools and

Page 45 - V3_3,V5_0, V12V

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 3 Plum A+3 Infusion SystemFigure 7-1. Bottom View7.2.4BATTERY, WIRE HARNESS, DOOR, AND DOOR PAD

Page 46 - BATTERY CHARGER

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 2 Technical Service Manual1.2CONVENTIONSThe conventions listed in Table 1-1 are used throughout thi

Page 47 - MECHANISMSUBSYSTEM

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 4 Technical Service ManualTo replace a battery, wire harness, door, or door pad, s

Page 48 - MOTOR POSITION SENSORS

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 5 Plum A+3 Infusion SystemFigure 7-2. AC Power Cord Assembly and Battery AssemblyBATTERY WITH W

Page 49 - V2_5 REFERENCE VOLTAGE

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 6 Technical Service Manual7.2.5AC POWER CORD, RETAINER, AND VELCRO STRAP REPLACEME

Page 50 - TransmitterCircuitry

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 7 Plum A+3 Infusion System7.2.6SEPARATING THE FRONT ENCLOSURE, REAR ENCLOSURE, AND MAIN CHASSIS

Page 51 - PRESSURE SENSORS

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 8 Technical Service ManualFigure 7-3. Rear Enclosure, Main Chassis, and Front Encl

Page 52 - AmplifierandLowPassFilter

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 9 Plum A+3 Infusion SystemFigure 7-4. Screw Placement Sequence7.2.7PERIPHERAL INTERFACE ASSEMBL

Page 53 - SERIAL EEPROM

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 10 Technical Service Manual6. Depress the retention clip and carefully pull the pe

Page 54 - PRINTEDWIRINGASSEMBLIES

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 11 Plum A+3 Infusion SystemFigure 7-5. Peripheral Interface Assembly and Peripheral PWAs7.2.9PE

Page 55 - PERIPHERALINTERFACEPWA

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 12 Technical Service Manual4. Using the X-acto knife, lift the volume control knob

Page 56 - DRIVERPWA

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 13 Plum A+3 Infusion System9. Using a 3/16 nutdriver, remove the screws from the DataPort conne

Page 57 - SWITCHPWA

1.4 ACRONYMS AND ABBREVIATIONSTechnical Service Manual 1 - 3 Plum A+3 Infusion System1.4ACRONYMS AND ABBREVIATIONSAcronyms and abbreviations used in t

Page 58 - MECHANICALOVERVIEW

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 14 Technical Service Manual7.2.10FRONT/REAR ENCLOSURE GASKET REPLACEMENTThe recomm

Page 59 - CASSETTE

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 15 Plum A+3 Infusion SystemFigure 7-7. Lower Front Enclosure Gaskets7.2.11.1EMI GASKET REPLACEM

Page 60

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 16 Technical Service Manual7.2.11.2KEYPAD GASKET REPLACEMENTThe recommended tool f

Page 61 - MECHANISMASSEMBLY

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 17 Plum A+3 Infusion System7.2.12REAR ENCLOSURE ASSEMBLY COMPONENT REPLACEMENTRear enclosure as

Page 62 - PLUNGER DRIVE SUBSYSTEM

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 18 Technical Service ManualFigure 7-9. Internal Rear Enclosure Assembly Components

Page 63 - MAINTENANCE AND SERVICE

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 19 Plum A+3 Infusion System7.2.12.1POLE CLAMP ASSEMBLY AND BACKING PLATE REPLACEMENTRecommended

Page 64 - Table 5-1. Cleaning Solutions

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 20 Technical Service ManualReplacement parts for this procedure are:Cord, Internal

Page 65 - PREVENTIVE MAINTENANCE

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 21 Plum A+3 Infusion System8. Join the rear enclosure and main chassis assembly in the exact re

Page 66 - LABELS INSPECTION

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 22 Technical Service Manual7.2.12.5REAR ENCLOSURE GASKET REPLACEMENTThe recommende

Page 67 - Figure 5-1. Infuser Labels

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 23 Plum A+3 Infusion System Figure 7-10. Minipole Assembly7.2.14MAIN CHASSIS ASSEMBLY COMPONENT

Page 68

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 4 Technical Service ManualLED Light emitting diodeL/S Line selectMB MegabyteMHz Megahertzmin Minute

Page 69

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 24 Technical Service ManualFigure 7-11. Main Chassis Assembly Components (1 of 2)I

Page 70 - OPEN PULL DOWN SPIN

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 25 Plum A+3 Infusion SystemFigure 7-11. Main Chassis Assembly Components (2 of 2)4-40 x 3/8 HEX

Page 71 - FLUID SHIELD INSPECTION

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 26 Technical Service Manual7.2.14.1POWER SUPPLY PWA REPLACEMENTThe recommended too

Page 72

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 27 Plum A+3 Infusion System6. Using the Phillips screwdriver, remove the screw that secures the

Page 73 - PRESSURE

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 28 Technical Service Manual7.2.14.4CPU/DRIVER CABLE REPLACEMENTThe recommended too

Page 74 - PROXIMAL

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 29 Plum A+3 Infusion SystemFigure 7-12. CPU/Driver Cable RoutingCENTER TABLOWER TABTOP TABFERRI

Page 75 - RUBBER FOOT PAD INSPECTION

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 30 Technical Service ManualFigure 7-13. Ferrite Tape Positioning (1 of 2)Figure 7-

Page 76 - Clamp slides to grip the pole

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 31 Plum A+3 Infusion System7.2.14.5MOTOR POWER CABLE REPLACEMENTThe recommended tool for this p

Page 77 - 5.2 PREVENTIVE MAINTENANCE

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 32 Technical Service Manual10. Disconnect the piezo alarm cable from the CPU PWA.1

Page 78 - Figure 5-12. Battery Doors

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 33 Plum A+3 Infusion System7.2.14.8MECHANISM ASSEMBLY REPLACEMENT Note: Replacing the mechanis

Page 79 - KEYPAD INSPECTION

1.5 USER QUALIFICATIONTechnical Service Manual 1 - 5 Plum A+3 Infusion System1.5USER QUALIFICATIONThe Plum A+3 must be used at the direction of or und

Page 80 - OFF (down) position

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 34 Technical Service Manual7.2.14.8.1ENTER THE SERIAL NUMBERAfter the mechanism as

Page 81 - Options/

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 35 Plum A+3 Infusion SystemTo verify successful mechanism chassis gasket tape installation, per

Page 82

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 36 Technical Service Manual6. Disengage the cassette door from the opener handle a

Page 83 - Line 1 Line 2 Line 3

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 37 Plum A+3 Infusion System14. Align the mechanism assembly pins, then install the replacement

Page 84 - PERFORMANCE VERIFICATION TEST

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 38 Technical Service ManualFigure 7-17. Close-up Views of Normal (left) and Bent (

Page 85 - TEST SETUPS

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 39 Plum A+3 Infusion SystemFigure 7-18. Fluid Shield ReplacementMECHANISMASSEMBLYFLUID SHIELDCA

Page 86 - BASIC TEST SETUP

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 40 Technical Service ManualFigure 7-19. Cassette Door and Opener Handle Assembly R

Page 87

7.2 REPLACEMENT PROCEDURESTechnical Service Manual 7 - 41 Plum A+3 Infusion System7.2.14.11OPENER HANDLE ASSEMBLY REPLACEMENTThe recommended tool for

Page 88 - Preparing the Primary Line

SECTION 7 REPLACEABLE PARTS AND REPAIRSPlum A+3 Infusion System 7 - 42 Technical Service ManualThis page intentionally left blank.

Page 89

Technical Service Manual 8 - 1 Plum A+3 Infusion SystemSection8SPECIFICATIONSThe following specifications apply to the Plum A+3 infusion system.PHYSI

Page 90 - Loading the Cassette

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 6 Technical Service Manual1.7.1UNPACKINGInspect the shipping container as detailed in Section 1.7.2

Page 91

SECTION 8 SPECIFICATIONSPlum A+3 Infusion System 8 - 2 Technical Service ManualENVIRONMENTOperating: 41 to 104 F (5 to 40 C); 10 % to 90 % relativ

Page 92 - Preparing the Secondary Line

Technical Service Manual 9 - 1 Plum A+3 Infusion SystemSection9DRAWINGSFigure 9-1 through Figure 9-10 show the Illustrated Parts Breakdown (IPB) and

Page 93 - CLAVE SECONDARY PORT

SECTION 9 DRAWINGSPlum A+3 Infusion System 9 - 2 Technical Service Manual7 PWA, Power Supply Section 7.2.14.18 Assembly, Mechanism Section 7.2.14.89 A

Page 94

SECTION 9 DRAWINGSTechnical Service Manual 9 - 3 Plum A+3 Infusion System37Assembly, MinipoleA: Hanger, BagB: Housing, ClutchC: Ring, CotterD: Spring,

Page 95

SECTION 9 DRAWINGSPlum A+3 Infusion System 9 - 4 Technical Service Manual65 Screw, 4-24 x 1/4, Pan Head, Phillips, SS As applicable66 Screw, 4-40 x 1/

Page 96

Technical Service Manual 9 - 5 Plum A+3 Infusion System12561415163418108919202425 2629 30323534271273133172321221113283737C37B37A37D36HOSPIRA, INC.Fig

Page 97 - DISTAL AIR-IN-LINE TEST SETUP

Plum A+3 Infusion System 9 - 6 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 98

Technical Service Manual 9 - 7 Plum A+3 Infusion System383940414447464551504952 53 545958575655656460726671697674817778 79 80 82 857583 848648626342 4

Page 99

Plum A+3 Infusion System 9 - 8 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 100

Technical Service Manual 9 - 9 Plum A+3 Infusion System17578717976787250324HOSPIRA, INC.Figure 9-2.Front Enclosures, Rear Enclosure,and Main Chassis A

Page 101 - DISTAL OCCLUSION TEST SETUP

1.7 INFUSION SYSTEM INSTALLATIONTechnical Service Manual 1 - 7 Plum A+3 Infusion System3. The LCD screen briefly displays the SELF TEST screen (see Fi

Page 102 - Setup Procedure

Plum A+3 Infusion System 9 - 10 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 103 - UNIVERSAL

Technical Service Manual 9 - 11 Plum A+3 Infusion System787120534948254350HOSPIRA, INC.Figure 9-3.Front Enclosure AssembliesDRAWING NO.NOT APPLICABLER

Page 104 - CASSETTE ALARM TEST

Plum A+3 Infusion System 9 - 12 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 105 - UNRESTRICTED FLOW TEST

Technical Service Manual 9 - 13 Plum A+3 Infusion System518482298482337716830828528265274HOSPIRA, INC.Figure 9-4.Rear Enclosure AssemblyDRAWING NO.NOT

Page 106

Plum A+3 Infusion System 9 - 14 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 107 - ALARM LOUDNESS TEST

Technical Service Manual 9 - 15 Plum A+3 Infusion System2759675758562078711HOSPIRA, INC.Figure 9-4.Rear Enclosure AssemblyDRAWING NO.NOT APPLICABLERev

Page 108 - KEYPAD LOCKOUT SWITCH TEST

Plum A+3 Infusion System 9 - 16 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 109 - PROXIMAL AIR-IN-LINE TEST

Technical Service Manual 9 - 17 Plum A+3 Infusion System12151434606786616880HOSPIRA, INC.Figure 9-5.Peripheral Interface AssemblyDRAWING NO.NOT APPLIC

Page 110 - DISTAL OCCLUSION TEST

Plum A+3 Infusion System 9 - 18 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 111

Technical Service Manual 9 - 19 Plum A+3 Infusion System713123938313246474041696618HOSPIRA, INC.Figure 9-6.Main Chassis AssemblyDRAWING NO.NOT APPLICA

Page 112

Plum A+3 Infusion System Technical Service ManualThis page intentionally left blank.

Page 113 - DELIVERY ACCURACY TEST

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 8 Technical Service Manual1.8BIOMED SETTINGSThe BIOMED SETTINGS screens contain the following opti

Page 114 - 12” - 24”

Plum A+3 Infusion System 9 - 20 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 115 - NURSE CALL TEST

Technical Service Manual 9 - 21 Plum A+3 Infusion System46651681973657017HOSPIRA, INC.Figure 9-6.Main Chassis AssemblyDRAWING NO.NOT APPLICABLERev. N/

Page 116 - ELECTRICAL SAFETY TEST

Plum A+3 Infusion System 9 - 22 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 117 - END OF THE PVT

Technical Service Manual 9 - 23 Plum A+3 Infusion System1870171640HOSPIRA, INC.Figure 9-7.CPU PWA, Display, and KeypadDRAWING NO.NOT APPLICABLERev. N/

Page 118

Plum A+3 Infusion System 9 - 24 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 119 - TROUBLESHOOTING

Technical Service Manual 9 - 25 Plum A+3 Infusion System4187766HOSPIRA, INC.Figure 9-8.CPU PWA and Main ChassisDRAWING NO.NOT APPLICABLERev. N/ASheet

Page 120 - OPERATIONALALARMMESSAGES

Plum A+3 Infusion System 9 - 26 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 121

Technical Service Manual 9 - 27 Plum A+3 Infusion System2523672442456444733743HOSPIRA, INC.Figure 9-9.AC Power Cord, Retainer, Batteries,and MinipoleD

Page 122 - SECTION 6 TROUBLESHOOTING

Plum A+3 Infusion System 9 - 28 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 123

Technical Service Manual 9 - 29 Plum A+3 Infusion System116486955910HOSPIRA, INC.Figure 9-10.Mechanism AssemblyDRAWING NO.NOT APPLICABLERev. N/ASheet

Page 124

1.8 BIOMED SETTINGSTechnical Service Manual 1 - 9 Plum A+3 Infusion SystemFigure 1-2. Biomed SettingsTable 1-2. System Configuration DataData Options

Page 125

Plum A+3 Infusion System 9 - 30 Technical Service ManualSECTION 9 DRAWINGSThis page intentionally left blank.

Page 126 - Replace piezo alarm

Technical Service Manual A - 1 Plum A+3 Infusion SystemAPPENDIXUSE OF THE INFUSION SYSTEM IN ELECTROMAGNETIC ENVIRONMENTSThe Plum A+3 Infusion System

Page 127

APPENDIXPlum A+3 Infusion System A - 2 Technical Service ManualELECTROMAGNETIC IMMUNITYTable A-2 details guidance for the electromagnetic environment.

Page 128

APPENDIXTechnical Service Manual A - 3 Plum A+3 Infusion SystemELECTROMAGNETIC IMMUNITY FOR LIFE-SUPPORTING EQUIPMENT AND SYSTEMSTable A-3 provides gu

Page 129

APPENDIXPlum A+3 Infusion System A - 4 Technical Service Manuala The industrial, scientific and medical (ISM) bands between 150 kHz and 80 MHzare 6.76

Page 130

APPENDIXTechnical Service Manual A - 5 Plum A+3 Infusion SystemRECOMMENDED SEPARATION DISTANCES FOR COMMUNICATIONS EQUIPMENTThe infusion system is int

Page 131

APPENDIXPlum A+3 Infusion System A - 6 Technical Service ManualThis page intentionally left blank.

Page 132

Technical Service Manual 430-95424-009-ENDFor technical assistance, product return authorization, and to order parts, accessories,or manuals within th

Page 133 - TROUBLESHOOTINGPROCEDURES

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 10 Technical Service Manual1.8.1IVPARAMETERSThe IV Parameters screen contains Common IV Parameters

Page 134

1.8 BIOMED SETTINGSTechnical Service Manual 1 - 11 Plum A+3 Infusion SystemFigure 1-4. Common IV ParametersFigure 1-5. Macro IV ParametersBIOMED SETTI

Page 135 - UNRESTRICTED FLOW

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 12 Technical Service Manual1.8.2ALARMSLOGThe Alarms Log will retain the latest 40 alarm and malfun

Page 136

1.8 BIOMED SETTINGSTechnical Service Manual 1 - 13 Plum A+3 Infusion System1.8.3SETTINGTHETIMEANDDATE Note: The infusion system will automaticall

Page 137 - AND REPAIRS

SECTION 1 INTRODUCTIONPlum A+3 Infusion System 1 - 14 Technical Service ManualThis page intentionally left blank.

Page 138 - RUBBER FOOT PAD REPLACEMENT

Technical Service Manual 2 - 1 Plum A+3 Infusion SystemSection2WARRANTYSubject to the terms and conditions herein, Hospira, Inc., hereinafter referre

Page 139 - REPLACEMENT

SECTION 2 WARRANTYPlum A+3 Infusion System 2 - 2 Technical Service ManualThis page intentionally left blank.

Page 140

Technical Service Manual 3 - 1 Plum A+3 Infusion SystemSection3SYSTEMOPERATINGMANUALA copy of the System Operating Manual is included with every Pl

Page 141 - 7.2 REPLACEMENT PROCEDURES

Technical Service Manual Plum A+3 Infusion SystemChangeHistoryPart Number Description of Change430-95424-008 (A, 2014-08)Eighth IssueUpdate front a

Page 142

SECTION 3 SYSTEM OPERATING MANUALPlum A+3 Infusion System 3 - 2 Technical Service ManualThis page intentionally left blank.

Page 143

Technical Service Manual 4 - 1 Plum A+3 Infusion SystemSection4THEORYOFOPERATIONThis section describes the Plum A+3 theory of operation. The theory

Page 144

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 2 Technical Service Manual4.2ELECTRONICSUBSYSTEMOVERVIEWThis section describes the functio

Page 145

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 3 Plum A+3 Infusion System4.2.1CPUSUBSYSTEMThe CPU subsystem contains the main microcon

Page 146 - PERIPHERAL PWA REPLACEMENT

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 4 Technical Service Manual4.2.1.3PROGRAMMABLE READ-ONLY MEMORYThe CPU subsystem has two 512

Page 147

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 5 Plum A+3 Infusion System4.2.1.7LCD BACKLIGHT CONTROLThe LCD panel is backlit by a cold

Page 148 - PERIPHERAL COVER REPLACEMENT

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 6 Technical Service Manual4.2.1.10VOLTAGE MONITOR WATCHDOG TIMERIt is important to protect t

Page 149

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 7 Plum A+3 Infusion SystemFigure 4-2. Serial Interface to ADCTable 4-1. Analog InputsSig

Page 150

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 8 Technical Service Manual4.2.1.12DIGITAL-TO-ANALOG CONVERTERThe dual 8 bit digital-to-analo

Page 151 - KEYPAD GASKET (3)

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 9 Plum A+3 Infusion System4.2.1.17NURSE CALL INTERFACEA nurse call relay switch on the p

Page 152 - TOP SEAL GASKET REPLACEMENT

CHANGE HISTORYPlum A+3 Infusion System Technical Service ManualThis page intentionally left blank.

Page 153

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 10 Technical Service ManualThe DataPort system conforms to the EIA-232-D standard, with the

Page 154

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 11 Plum A+3 Infusion SystemLegend: P = Power A = Analog D = Digital I = Input O = Outpu

Page 155

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 12 Technical Service ManualLegend: P = Power A = Analog D = Digital I = Input O = OutputFLP

Page 156 - AC CONNECTOR REPLACEMENT

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 13 Plum A+3 Infusion System4.2.2POWERSUPPLYSUBSYSTEMThe power supply subsystem provide

Page 157 - FUSE REPLACEMENT

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 14 Technical Service Manual4.2.2.1.3CurrentLimitLoopThe current limit loop is activated wh

Page 158 - MINIPOLE ASSEMBLY REPLACEMENT

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 15 Plum A+3 Infusion SystemThe infuser is activated when the [ON/OFF] key is pressed or

Page 159

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 16 Technical Service Manual4.2.2.8BATTERY CHARGERThe software battery management algorithm c

Page 160

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 17 Plum A+3 Infusion System4.2.3MECHANISMSUBSYSTEMThe mechanism subsystem includes the

Page 161

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 18 Technical Service ManualThe unipolar motor windings have a center tap connected on each o

Page 162 - KEYPAD REPLACEMENT

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 19 Plum A+3 Infusion SystemEach optical switch consists of an infrared LED that shines t

Page 163 - DISPLAY ASSEMBLY REPLACEMENT

Technical Service Manual iii Plum A+3 Infusion SystemContentsSECTION 1INTRODUCTION . . . . . . . . . . . . .

Page 164 - CPU/DRIVER CABLE REPLACEMENT

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 20 Technical Service Manual4.2.3.4AIR SENSORSThe mechanism subsystem includes two air sensor

Page 165 - LOWER TAB

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 21 Plum A+3 Infusion System4.2.3.4.2ReceiverCircuitryWhen the cassette’s test port is f

Page 166 - INSULATION TAPE

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 22 Technical Service ManualFigure 4-6. Pressure Sensor Excitation and Amplifier Block Diagra

Page 167 - CPU PWA REPLACEMENT

4.2 ELECTRONIC SUBSYSTEM OVERVIEWTechnical Service Manual 4 - 23 Plum A+3 Infusion System4.2.3.6PRESSURE SENSOR CALIBRATIONPressure sensors are calibr

Page 168

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 24 Technical Service Manual4.3PRINTEDWIRINGASSEMBLIESInfusion system electronics are packa

Page 169

4.3 PRINTED WIRING ASSEMBLIESTechnical Service Manual 4 - 25 Plum A+3 Infusion System4.3.2PERIPHERALPWAThe peripheral PWA contains part of the CPU su

Page 170 - ENTER THE SERIAL NUMBER

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 26 Technical Service Manual4.3.4CPUPWAThe CPU PWA contains most of the CPU subsystem functi

Page 171 - LOCATION

4.3 PRINTED WIRING ASSEMBLIESTechnical Service Manual 4 - 27 Plum A+3 Infusion SystemSee Table 4-9 for driver PWA interface connections.4.3.6SWITCHPW

Page 172

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 28 Technical Service Manual4.4REMOTEMOUNTEDPERIPHERALSRemote mounted peripherals include t

Page 173

4.5 MECHANICAL OVERVIEWTechnical Service Manual 4 - 29 Plum A+3 Infusion System4.5.1CASSETTEThe cassette (see Figure 4-7 and Figure 4-8) operates on a

Page 174 - Bent TabNormal Tab

CONTENTSPlum A+3 Infusion SystemivTechnical Service Manual4.2.2 POWER SUPPLY SUBSYSTEM . . . . . . . . . . . . .

Page 175 - GASKET, .72 IN. (2)

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 30 Technical Service ManualFigure 4-7. Major Elements of the Dual-Channel CassetteFigure 4-8

Page 176

4.5 MECHANICAL OVERVIEWTechnical Service Manual 4 - 31 Plum A+3 Infusion System4.5.2MECHANISMASSEMBLYThe mechanism assembly is a fully self-contained

Page 177

SECTION 4 THEORY OF OPERATIONPlum A+3 Infusion System 4 - 32 Technical Service ManualFigure 4-9. Mechanism Valve Pins and Sensor Locations4.5.2.3INLET

Page 178

Technical Service Manual 5 - 1 Plum A+3 Infusion SystemSection5MAINTENANCE AND SERVICE TESTSA complete maintenance program promotes infusion system l

Page 179 - SPECIFICATIONS

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 2 Technical Service Manual Note: At the time of publication, Hospira recommends o

Page 180 - SECTION 8 SPECIFICATIONS

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 3 Plum A+3 Infusion System5.2PREVENTIVE MAINTENANCEThe Plum A+3 is three infusers in a single h

Page 181 - DRAWINGS

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 4 Technical Service Manual5.2.1LABELS INSPECTIONVisually inspect the infuser label

Page 182 - SECTION 9 DRAWINGS

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 5 Plum A+3 Infusion SystemFigure 5-1. Infuser LabelsREAR CAUTIONLABELPOLE CLAMPLABELCLAMP WARNI

Page 183

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 6 Technical Service Manual5.2.2AC POWER CORD, RETAINER, AND VELCRO STRAP INSPECTIO

Page 184

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 7 Plum A+3 Infusion System5.2.3FRONT ENCLOSURE AND REAR ENCLOSURE INSPECTIONVisually inspect th

Page 185

CONTENTSTechnical Service ManualvPlum A+3 Infusion System5.2.12 DISPLAY AND INDICATORS INSPECTION . . . . . . . . . . .

Page 186

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 8 Technical Service Manual5.2.4DOOR LEVER ASSEMBLY, DOOR ROLLER,DOOR INSPECTION, A

Page 187

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 9 Plum A+3 Infusion System5.2.5FLUID SHIELD INSPECTIONVisually inspect the Plum A+3 fluid shiel

Page 188

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 10 Technical Service ManualFigure 5-6. Inspecting the Fluid Shield with Feeler Gau

Page 189

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 11 Plum A+3 Infusion System5.2.6DISTAL PRESSURE PIN INSPECTIONVisually inspect the distal press

Page 190

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 12 Technical Service Manual5.2.7PROXIMAL PRESSURE PIN INSPECTIONVisually inspect t

Page 191

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 13 Plum A+3 Infusion System5.2.8RUBBER FOOT PAD INSPECTIONPerform a visual inspection of the ru

Page 192

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 14 Technical Service Manual Note: The A+3 infuser utilizes a heavy duty clamp tha

Page 193

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 15 Plum A+3 Infusion System3. Carefully pull the battery and wire harness assembly out of the e

Page 194

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 16 Technical Service ManualFigure 5-12. Battery DoorsFigure 5-13. Removing the Bat

Page 195

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 17 Plum A+3 Infusion System5.2.11KEYPAD INSPECTIONVisually inspect the keypads at least once ev

Page 196

CONTENTSPlum A+3 Infusion SystemviTechnical Service Manual7.2.10 FRONT/REAR ENCLOSURE GASKET REPLACEMENT . . . . . . . 7-147.2.1

Page 197

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 18 Technical Service Manual5.2.12DISPLAY AND INDICATORS INSPECTIONVisually inspect

Page 198

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 19 Plum A+3 Infusion SystemFigure 5-15. Display and IndicatorsBackPrimeOptions/Vol InfABLINE FL

Page 199

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 20 Technical Service Manual5.2.13KEYPAD LOCKOUT SWITCH INSPECTIONInspect the keypa

Page 200

5.2 PREVENTIVE MAINTENANCETechnical Service Manual 5 - 21 Plum A+3 Infusion SystemPlum A+3 Infusion SystemPreventive Maintenance ChecklistLine 1 Line

Page 201

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 22 Technical Service Manual5.3PERFORMANCE VERIFICATION TESTThe Performance Verific

Page 202

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 23 Plum A+3 Infusion System5.3.1PVT EQUIPMENT LISTTo complete all the performance verifi

Page 203

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 24 Technical Service Manual5.3.2.1BASIC TEST SETUPThe Basic test setup consists of

Page 204

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 25 Plum A+3 Infusion SystemThe following tests can be performed using one Basic test set

Page 205

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 26 Technical Service Manual5.3.2.1.2Preparing the Primary LineTo prepare the prima

Page 206

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 27 Plum A+3 Infusion System5. Suspend the container on an IV pole.6. Check for leaks. If

Page 207

Technical Service Manual vii Plum A+3 Infusion SystemFiguresCONTENTSFigure 1-1 Display and Keypad . . . . . . . . . . . . . . . . . . . . . . . . .

Page 208

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 28 Technical Service Manual10. Tap and clear air from the cassette, Y-site, and tu

Page 209

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 29 Plum A+3 Infusion System2. Grasp the cassette by the finger grip (seeFigure5-27).Figu

Page 210

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 30 Technical Service Manual5.3.2.1.4Preparing the Secondary LineTo prepare the sec

Page 211 - APPENDIX

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 31 Plum A+3 Infusion System5. Slowly open the roller clamp to allow fluid to flow into t

Page 212 - ELECTROMAGNETIC IMMUNITY

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 32 Technical Service Manual• If the cassette has a capped secondary port: Confirm

Page 213

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 33 Plum A+3 Infusion SystemThe test setup is complete, as shown in Figure 5-35. Figure 5

Page 214

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 34 Technical Service Manual5.3.2.2PROXIMAL AIR-IN-LINE TEST SETUPThis section desc

Page 215 - FOR COMMUNICATIONS EQUIPMENT

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 35 Plum A+3 Infusion System5.3.2.3DISTAL AIR-IN-LINE TEST SETUPThis section describes th

Page 216

SECTION 5 MAINTENANCE AND SERVICE TESTSPlum A+3 Infusion System 5 - 36 Technical Service Manual5.3.2.4PRIMING A RUN-IN CASSETTE ASSEMBLYPrimed run-in

Page 217 - MCN 160992

5.3 PERFORMANCE VERIFICATION TESTTechnical Service Manual 5 - 37 Plum A+3 Infusion System5. Press and hold [BACKPRIME] to pump water from the drip cha